Page 47: Flashing 9s flashing 0s

Advertising
background image

To delete any calibration data entered by the user:

1. Turn off the instrument.
2. Push and hold CAL/Zero.
3. Turn on the instrument.

The CAL? light flashes. The instrument starts in Calibration mode.

4. Calibrate the instrument before use.

Flashing 9s

When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument. The
display will also flash 9s if Ratio is off and the measurement is greater
than 40 NTUs (1000 FNUs or 9.8 EBCs). Turn Ratio on. Refer to

Measure over-range samples

on page 24.

Flashing 0s

When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument (i.e., less than
20 FAU) or a negative value. Calibrate the instrument.

• When measuring absorbance or transmittance, set the zero reference

point again.

• When measuring absorbance, make sure that the reading is positive.

To measure samples with negative absorbance, set the analytical
zero using the sample with the greatest absorbance and read the
sample with the least absorbance. Record the reading as negative
absorbance.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Replacement parts

Description

Quantity

Item no.

Calibration kit, StablCal

®

, sealed sample

cells (<0.1, 20, 200, 1000, 4000 and
7500 NTU)

1

2659505

Cover, sample cell compartment

1

4702500

Cover, printer

1

4705400

Dust cover

1

4703000

Fuse for 115 V operation, time-delay,
250 V, 1.6 A, UL/CSA approved

1

3030700

Fuse for 230 V operation, time-delay,
250 V, 1.6 A, IEC type, VDE approved

1

3030600

Gelex

®

secondary turbidity standardization

kit
(stray light standard and 0–2, 0–20, 0–200,
200-4000, and 4,000–10,000 NTU)

1

2589200

Oiling cloth

1

4707600

Power cord, North America, 115 VAC,
UL/CSA approved

1

1801000

Power cord, European, 230 VAC, VDE
approved

1

4683600

Printer paper, thermal

5

4709000

Sample cells, 30mL, 1 in. round glass

6

2084900

Silicone oil

1

126936

English 45