English, Specifications – Hach-Lange BIOGAS Basic User Manual User Manual

Page 39

Advertising
background image

English 39

English

Specifications

Specifications are subject to change without notice.

Methods

End point titration

1 to 3 pre-set

Inflection point titration

Auto determination of 1 to 4 inflection points with pro-
grammable IP acceptation windows.
IP detection using 1st and 2nd derivative curve.

Titration stops at

pH, mV, mL, IP number

Titrant addition techniques

Incremental dynamic, incremental monotonic and con-
tinuous dynamic.

Titrant calibration

pH electrode calibration

Up to 5 points using IUPAC standards or 4-7-10 Series
buffers with error recognition on buffer used.

Direct pH/mV measurements

With recording on stable reading.

Back titration with manual
reagent addition
Sequencing of up to 3 methods
Coupling of 2 methods in one beaker

Measuring ranges - Resolution

pH value

-9 to 23 pH - 0.001 pH

mV value

±2000 mV - 0.1 mV

Temperature

–10 °C to +100 °C (14 to 212 °F) - 0.1 °C

Printout

Automatic. GLP compliant
3 levels of detail

Results

In each method, calculation of up to 4 results and 2
user-defined equations

Units

All standard units for samples/results
User-defined result units

Storage capacity

Global password protection for programming access
Non-volatile memory for storage of the last 60 results
10 user programmable methods
Libraries with more than 30 electrodes and 20 titrants
pre-identified
Embedded operating procedures for burette and
reagent exchange operations

Sample list

Up to 20 data with alphanumeric ID

Electrode stand - stirring

Magnetic stirrer

22 reproducible speeds (0 to 1100 rpm) in 50 rpm steps

Propeller connection
Beaker volume

5 to 400 mL