Run analyses – Hach-Lange BIOGAS Basic User Manual User Manual

Page 52

Advertising
background image

English 52

Cell menu

Press three times the RIGHT arrow key in the Main menu to move to the Electrode menu.

Figure 8 Cell menu

Apply internal stirring

1.

Select Internal stirring = ON,

2.

In the field Speed, select a stirring speed.

Apply external stirring

1.

Connect the Stirring Propeller, part no. 904-480, to the Titration Manager (see

Figure 2 on page

45

, position 14).

Line 1 is automatically replaced by External stirring.

2.

Select External stirring = ON.

3.

Adjust stirring by turning the stirring propeller knob.

Note:

You can consult the corresponding table between the position (1 to 9) and the stirring

speed by selecting the field Speed setting.

Run analyses

Run an elecrode calibration

Refer to Short-Form Reminder no. 4.
1.

Select the method.

2.

Install the electrode group.

3.

Enter the Electrode menu.

4.

Select 1 Calibrate electrodes.

5.

Press the validation key.

6.

Select the electrode to be calibrated from the proposed list.

7.

Press 1 to start the calibration.

1

Title bar: indicates the name of the window.

3

Speed: select the internal stirring speed, from
100 to 1100 rpm by steps of 50 rpm.

2

Internal stirring: command stirrer On/Off (see

Apply internal stirring

).

4

Cell icon: animated icon indicates when the mag-
netic stirrer or propeller is operating.

1

3

2

4

Advertising