Hach-Lange BIOGAS Basic User Manual User Manual

Page 51

Advertising
background image

English 51

Electrode menu

Press twice the RIGHT arrow key in the Main menu to move to the Electrode menu.

Figure 7 Electrode menu

1

Title bar: indicates the name of the window.

4

GLP-Archives: access GLP tables and visualise
the last electrode calibration results.

2

Calibrate electrode: run a calibration or a cali-
bration sequence using the installed electrodes.

5

Electrode library- Supervisor use only: create,
edit, reset and delete electrodes stored in the
instrument.

3

Display measurement: displays mV and/or pH
and/or temperature at a connected electrode of
the electrode system.

6

Electrode icon: Electrode status icon indicates
the state of the electrode system. Four types of
icons can be displayed. See table on the next
page.

Table 2 Electrode icons

Icon

Description

Sunny icon:
The calibration has been performed on the electrode present in the system. Everything
is just right!

Cloudy icon:
The electrode calibration of the electrode present in the system should be performed
within 12 or 24 hours.

Stormy icon:
The calibration date has elapsed for the electrode present in the system.
If acceptance limits have been set for the calibration: at least one calibration result lies
outside the programmed acceptance limits.

Question mark:
There is a problem in the editing of the electrode system. You need to be in Supervisor
mode to solve the problem. Check the sequence/method parameters of the reagent or
the electrode.
Check the sequence or method, (press 1 in the Main window). The instrument indicates
the possible errors and prompts you to correct them until ? disappears.

1

3

2

4

5

6

Advertising