Maintenance recommendations, Form no. 9 – Hach-Lange TITRALAB Short-form Reminders User Manual

Page 19

Advertising
background image

Maintenance recommendations

Daily

Check and maintain electrodes according to manufacturer recommendations.

Check that there is no crystallisation at piston level, if so rinse the burette using
appropriate solvent to remove crystals. If not sufficient, dismount burette to remove
crystals.

Check that there is no crystallisation at addition tip end that may block the reagent to
pour freely, if so use appropriate solvent to remove crystals.

Check the level of reagent in the reagent bottle

Empty waste bottle if installed.

Remove air bubbles from the burette and tubing by running a “Flush” procedure from
the burette menu.

Weekly

Flush the burette with appropriate solvent to
ensure complete cleaning of burette and
stopcock.

Check the piston for leakages.

The burette stand must be changed when

liquid can be seen between the two sealing
lips. Refer to diagram.

piston

Reagent

glass

cylinder

Change burette

stand

Reagent

For burette air bubble removal,

run the Flush procedure

Form no. 9

Advertising