English, Specifications – Hach-Lange HT 200 S User Manual

Page 93

Advertising
background image

English 93

English

Specifications

These are subject to change without notice.

HT 200 S

Type

Fully automated high-temperature thermostat:
1000 watt high-performance heating
Fast cooling through two ventilator system

Digestion temperatures

40–150 °C or 170 °C

Digestion vessel compartments

Twelve digestion compartments for HachLange round cuvettes and reaction tubes

Digestion times

5–240 minutes

Heating-up time

Maximum 8 minutes

Cooling-down time

Maximum 13 minutes at an ambient temperature of < 25 °C

Operating modes

Three standard programs
Nine possible user-specific programs

Display

Two-line alphanumeric display

Temperature accuracy

Thermostat block ± 3 °C

Power consumption

1300 VA

Supply voltage

230 V
50 Hz

Cold instrument socket fuse

T 6.3A L; 250 V

Dimensions

Width 300 mm/height 330 mm/depth 430 mm (width 11.81 in./height 12.99 in./depth 16.93 in.)

Weight

12 kg (26.45 lb)