Laurel Electronics LAUREATE SERIES 2 SERIAL INPUT METER & REMOTE DISPLAY User Manual

Page 25

Advertising
background image

25

Input+Scaling tab. Works only with the Custom ASCII protocol, not the Modbus protocol,
since a meter cannot listen to commands while in the Rmt Dsply C display type.

Executing a Main Menu > Counter > Put Setup command downloads the setup information
into the meter, including the Rmt Dsply C display mode if selected. Do not execute a Main
Menu > Counter > Get Setup command if you have placed the meter into the Rmt Dsply C
mode, since normal meter communications cannot occur in that mode. Our Instrument
Setup Software will then reset the meter to Rmt Dsply A so that it can retrieve the setup
information from the meter.

For more information on serial communications, please download our Series 2 Serial Com-
munications Manual (for Custom ASCII Protocol) and our Series 2 Modbus Communications
Manual (for Modbus Protocol).

For Control Inputs, select the desired action of grounding Control Input 1, Control Input 2,
and both Control Inputs 1 and 2.

• Meter Reset is a master reset like removing and reapplying power.

• Function Reset resets latched alarms.

• Alarm Active allows external control of relays, as opposed to using normal setpoint

operation.

• Display Blanking blanks the meter display.

• Enable Tare enables a meter tare function and enters the current reading as a value to

be subtracted as a tare value.

• Tare if Enabled subtracts the tare value from the current reading.

Ignore the following selections, which do not apply to remote display operation: Peak,
Valley, External Gate, Meter Hold, Total A Reset, Total B Reset.