Instructions, General workflow coupling protocol – Luminex xMAP Antibody Coupling Kit User Manual

Page 12

Advertising
background image

Page 8 of 20

Instructions

General Workflow

Coupling Protocol

The following is the detailed step-by-step process for coupling antibodies (or similar
proteins) to carboxylated magnetic microspheres. An abridged quick reference guide
(part number 89-30000-00-361) can be found online at http://www.luminexcorp.com/
downloads.

IMPORTANT: Protect photosensitive microspheres from light whenever possible
throughout this entire procedure.

1.

Remove kit and all reagents from the refrigerator and allow them to equilibrate to
room temperature for 20-30 minutes.

2.

While the reagents equilibrate to room temperature, calculate and note the required
volumes for Decision Point steps 4, 8, 11, 13, 19, and 20; for each reaction being
performed.

3.

Resuspend the stock microspheres.

If using a 1mL stock microsphere vial, vortex the stock microsphere vial for 10
seconds and then sonicate for 10 seconds to disperse the microspheres.
Alternatively, the microsphere vials can be rotated on a rotator for 15 minutes.

If using a 4mL stock microsphere vial, rotate the vial for 15 minutes at 15-30
rpm.

Advertising