Luminex xMAP Antibody Coupling Kit User Manual

Page 19

Advertising
background image

Page 15 of 20

4.

Prepare a solution of phycoerythrin-labeled anti-species IgG detection antibody at 4
μg/mL in PBS-1% BSA. Prepare a 1:2 dilution series of that detection antibody
solution to a concentration of 0.0625 μg/mL as shown in the following table.

NOTE: For optimal results use Costar round-bottom 96-well plates and Luminex
Magnetic Plate Separator with MagPlex microspheres.

5.

Aliquot 50 μL of the microsphere solution prepared in Step 3 into two entire columns
of wells of the plate (duplicate sets of 8 wells, 16 wells total).

6.

Add 50 μL of PBS-1% BSA, as a blank sample, into the wells in Row A containing
the microsphere solution.

7.

Add 50 μL of each of the diluted detection antibody solutions prepared in Step 4 into
the appropriate wells of the plate (as shown in the plate layout below).

1

2

3

4

5

6

7

8

9

10

11

12

A

Blank

Blank

B

1:64

1:64

C

1:32

1:32

D

1:16

1:16

E

1:8

1:8

F

1:4

1:4

G

1:2

1:2

H

1:1

1:1

Example plate layout using columns 1 & 2

Dilution Tube

Volume of

PBS-1% BSA

Volume of Detection

Antibody

Concentration

1:1

-

-

4 μg/mL

1:2

500 μL

500 μL from Tube 1:1

2 μg/mL

1:4

500 μL

500 μL from Tube 1:2

1 μg/mL

1:8

500 μL

500 μL from Tube 1:4

0.5 μg/mL

1:16

500 μL

500 μL from Tube 1:8

0.25 μg/mL

1:32

500 μL

500 μL from Tube 1:16

0.125 μg/mL

1:64

500 μL

500 μL from Tube 1:32

0.0625 μg/mL