Performance specifications, Speed, Accuracy and precision – Luminex 100 IS Version 2.1 User Manual

Page 8: Sensitivity

Advertising
background image

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP Technology

2

PN 89-00002-00-062 Rev. A

Performance

Specifications

Speed

•

Speed: greater than or equal to 933 MHz Pentium III processor
with 256 MB RAM

•

USB cabling for fast data transfer

•

Automatic transfer of assay templates and new reagent
information into the system using a 3½” inch floppy disk or large
capacity read/write CD

•

Installation < 4 hours

•

System calibration < 10 minutes

•

System controls < 10 minutes

•

Barcode reader entry of sample IDs

•

Automatic post analysis

•

Analyze one 96-well plate/hour depending on manufacturer’s kit

•

Up to 100 xMAP microsphere sets per sample

•

Sheath flowrate 90

µL/sec. ± 5 µL

•

Sample injection rate into detector area 1

µL/sec. ± 0.05 µL

•

System warmup 30 minutes: Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.

Accuracy and Precision

•

Sample uptake volume accuracy ± 5%

•

Classification of xMAP microspheres > 80.0%

•

Misclassification of xMAP microspheres < 0.5%

•

Temperature control 0°C to ± 2°C of target

•

Internal sample carry over < 0.9%

•

Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity value

Sensitivity

•

Detect 1000 fluorochromes Phycoerythrin (PE) per microsphere

•

Reporter channel dynamic range: 3.5 decades