Capacity – Luminex 100 IS Version 2.1 User Manual

Page 9

Advertising
background image

x

MAP Technology

Luminex xMAP Technology

PN 89-00002-00-062 Rev. A

3

Capacity

The following specifications reflect minimum capacity values:
•

10 GB hard drive

•

Store data for up to 240,000 test results

•

1.4 MB 3½” inch floppy disk

•

100 MB read/write CD ROM

•

Analyze multiple 96-well plates per batch

•

Analyze multiple assay templates per plate

•

Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample

•

Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1 - 100 unique xMAP microsphere sets
in a single sample

•

Reporter channel range: 3.5 decades of detection

•

Collect from 10 to 100,000 events per xMAP microsphere set
specified

•

2.5 x 10

6

microspheres/mL maximum bead concentration

•

Sample uptake volumes from 20 to 200

µL

•

Sample core: 15-20 µm core at 1 µL/sec. sample injection rate

•

Sample timeout limit from 1 to 400 seconds

•

Automatic sampling from a 96-well plate

•

Start sampling from any well position

•

Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills

•

96-well microtiter plates types:

Must be compatible with XYP instrument plate holder: Flat
bottom, conical, round, filter bottom, half-well plates with
overall height no more than 0.75 inches (19 mm). Use black
opaque plates if possible to reduce photobleaching of beads.

•

Microtiter plates with 96 wells must be compatible with XYP
instrument heater block temperature from 35°C to 55°C
(tolerance 0°C to ± 2°C) (95°F to 131°F) when performing
heated assays and using the heater block