Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use

7

9. A dialog box opens with two installation options, Complete and

Custom. Select Complete and click Next to continue.

10. Click Install at the next dialog box to continue installation, which

may take several minutes.

11. Click Finish at the final InstallShield Wizard to complete the

installation.

12. Double-click the Luminex xPONENT icon on your desktop.
13. A license agreement dialog box opens. Click I accept the terms in

this license agreement, and click OK to continue.

14. A new license dialog box opens displaying the licenses for the

installation. Click OK to complete the installation and open the
xPONENT 3.1 SP1 login screen.

To install the xCONVERT Data Conversion Utility:

1. Click Install xPONENT file conversion utility. A File Down-

load – Security Warning dialog box opens.

2. Click Run. An Internet Explorer – Security Warning dialog

box opens.

3. Click Run. An xCONVERT Setup dialog box opens (.NET

Framework 2.0).

Note: Steps 3 and 4 may not
occur in all installations.

4. Click Accept. An xCONVERT Setup dialog box opens (for

SQL Server 2005 Express Edition).

5. Click Accept. An xCONVERT Setup dialog box opens (Visual

C++ Runtime Libraries (x86).

6. Click Install. A series of installation dialog boxes open as the

software installs. This may take several minutes.

7. The xCONVERT Setup Wizard opens. Click Next to continue.
8. Select an installation folder for the software. Click Next to

continue.

9. The Confirm Installation dialog box opens. Click Next to

continue.

10. An Installation Complete dialog box opens. Click Close to

complete the installation

.

Upgrade Firmware

The Luminex SD system, analyzer, and XYP instrument use
firmware to perform internal control functions. You must perform a
firmware upgrade for each of these instruments when you upgrade
the software. You must upgrade the firmware in the following order:

SD (if present)

LX100/200

XYP

Note: The xPONENT 3.1 SP1
CD installs a 90 day trial version
of the base package software.
Contact Luminex Technical
Support for a license key for the
full base version. See "Obtain
License Key" on page 12.

Note: Please see “Licensing” in
the “Using” chapter of the
xPONENT 3.1 software manual
if you need information on how
to change licensing information.

Advertising