Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use

4

These license packages are part of the upgrade kit only if you
purchase them from Luminex:

21 CFR Part 11 package

LIS package

Automation package

Security package

Extra seats package

Computer and

Software Upgrade

Instructions

You must run Luminex xPONENT 3.1 SP1 on a PC which meets the
minimum specifications listed in "System Requirements" on page 2.
To upgrade your PC, contact Luminex Technical Support. See
"Luminex Contact Information" on page 12.

If you are installing a new PC, continue with "Install New PC"
on page 4
.

If you are not installing a new PC, continue with "Upgrade
Firmware" on page 7
.

Install New PC

To install the new PC:

1. Close all applications. Perform a complete system shutdown.

Turn the power off and unplug the Luminex analyzer, Luminex
XYP, and PC.

2. Remove all connections from the old PC. Remove the old PC

and set it aside. Set the new PC in place.

3. Connect the keyboard and mouse to an available USB port, and

bar code reader to an available USB or PS2 port.

4. Connect the Luminex XYP serial cable to the back of the XYP

and the computer serial port.

5. Do NOT connect the Luminex 100/200 serial cable at this time.

You must update the firmware before connecting the Luminex
100 serial cable.

6. Connect the monitor cable to the computer. Touchscreen

monitors will have two cables.

7. Connect power cables for the monitor, computer, Luminex

analyzer, and Luminex XYP.

8. If you are installing a touchscreen monitor, you will need to

install the touchscreen drivers and calibrate the touchscreen
according to manufacturer’s instructions, or contact Luminex
Technical Support for assistance.

Note: If you did not get a new
PC with your xPONENT 3.1
SP1 upgrade, skip this step.