Import cal or ver kits, Export cal or ver kits, Delete cal and ver kit information – Luminex xPONENT 3.1 Rev 2 (IVD) User Manual

Page 87: Create calibration reports

Advertising
background image

WARNING:

Inaccurate data entry can invalidate the results of these
procedures. Careful entry of data in each field is necessary to
ensure accurate results.

1. On the Home page, click System Initialization.

2. Click Import Kit at the bottom right side of the window. The Import Calibration or

Performance Kit dialog box opens.

3. Locate the xMAP calibrator or verification kit information file (.lxl) and click Open.

Import CAL or VER Kits

1. Open the Maintenance page, then open the Lot Management tab.

2. Click Import Kit. The Import Calibration or Performance Kit dialog box opens.

3. Select the kit you wish to import, then click Open.

Export CAL or VER Kits

1. Open the Maintenance page, then open the Lot Management tab.

2. In the Active Reagents section, select the kit you want to export from the Calibration

Kit or Performance Verification Kit lists.

3. Click Export.

4. In the Export Calibrator or Verification Lot dialog box, select an export location for the

kit.

5. Type the kit name for the exported kit into the File Name box.

6. Click Save. The dialog box closes.

Delete CAL and VER Kit Information

1. Open the Maintenance page, then open the Lot Management tab.

2. In the Active Reagents section, select the kit you want to delete from the Calibration

Kit or Performance Verification Kit lists.

3. Click Delete Kit.

CAUTION:

There is no confirmation dialog box when you delete a kit.

Create Calibration Reports

1. Open the Results page, then open the Reports tab.

2. In the Report list, click Calibration Summary.

3. In the Type list, click the calibrator or verifier for which you want to create a report.

4. In the Dates list, select a start date for the date range you want to view.

5. In the Through list, select the end date for the date range you want to view.

6. Click Generate to display the desired report.

7. Use the left or right Page arrows to navigate to the different report pages.

For In Vitro Diagnostic Use

Using the Software

77

Advertising