Meriam 4150X ATEX I.S. Certified HART Communicator User Manual

Page 23

Advertising
background image

19

The editing function allows review of all HART parameters in a config-
uration and modification of editable parameters. The edit function can be
locked out if desired (see the Lockout section of this manual for more
information). To view detailed information of a stored configuration, move
the cursor to the desired tag number and press the Select soft key. The
Config Detail screen (see example at right) provides information to help
the user confirm the identity, origin, and save date/time of the configuration.
Pressing Select again opens the configuration for review or offline editing.

New configurations saved are added to the bottom of the List / Edit Config
list. If multiple entries for one device are shown, the most recently saved
configuration is always shown at the bottom.

Entries in the List / Edit Config list may have been saved or created from
various sources. Possible sources include the save command from a
connected HART device, created in Offline Mode using the Create Configs
function, copied from another configuration and renamed, or created by editing an existing file. The unit
will also list configurations downloaded from the separately available Device Management System software
from Meriam (see www.meriam.com/dm_solns.htm for more information). Tag numbers, date / time of
save, type of configuration stored, and other information are displayed under List/Edit Configs.
Configuration types are:

Complete Config = saved from connected device in Online mode, or a
complete config that was downloaded from DMS
Default Config

= created by 4150 / 4150X in Offline mode, or created in DMS

and downloaded to unit
Edit Vars Only

= created by editing an existing configuration and includes only the edited parameters

MVar Primary Cfg = for multivariable devices – Primary Variable configuration
MVar Sub Config = for multivariable devices – Sub-configuration (secondary, tertiary, fourth, etc
variables)
3 Create Configs:
The Create Configs function allows the user to configure a HART device file in Offline Mode for sending
to the intended device later when connected in the Online Mode. Use the soft key controls to move through
the list and select the device model number required. The menus prompt the user to make the necessary
configuration selections, tag the file for later retrieval and use, and edit configuration lines as required. This
function can be locked out. See the Lockout Section of this manual for more information.

4 Delete All Configs:
The Delete All Configs function clears all configurations from memory (including active and hidden
configs). When this menu option is selected, the unit will ask “Delete ALL Configs… ARE YOU
SURE???” Select Yes to clear all stored configurations from memory.

Delete Individual Configs / Clearing Configuration Memory:

Individually stored Configurations may be deleted with Del soft key function
(see example at right). When this menu option is selected, the unit will ask
“Delete CURRENT Config… ARE YOU SURE???” Select Yes to clear only
the selected configuration from memory.

The Del key only hides the record from the display screen; it does not clear
memory space. This function can be locked out. See the Lockout Section of
this manual for more information.

Offline

#

List/Edit Configs
Tag: PT-105A

Device ID 1365
PV: Pressure

Endress + Hauser
Cerabar S
Complete Config
Uses DOF 11070702
51 Variables
12/18/03 4:58 PM

Copy Del Select Back

Offline List / Edit

Configs Detail Screen

Offline

#

List/Edit Configs
Tag: PT-105A

Device ID 1365
PV: Pressure

Endress + Hauser
Cerabar S
Complete Config
Uses DOF 11070702
51 Variables
12/18/03 4:58 PM

Copy Del Select Back

Offline List / Edit

Configs Detail Screen

Advertising