Exchanging the purge valve frit or the purge valve, Solvent cabinet – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 108

108
Repairing the Pump
Exchanging the Purge Valve Frit or the Purge Valve
Exchanging the Purge Valve Frit or the
Purge Valve
1
Using a 1/4 inch wrench disconnect the pump outlet capillary at the purge
valve.
2
Disconnect the waste tube. Beware of leaking solvents due to hydrostatic
pressure.
3
Using the 14 mm wrench unscrew the purge valve and remove it.
4
Remove the plastic cap with the gold seal from the purge valve.
5
Using a pair of tweezers or a toothpick remove the frit
Figure 17
Purge Valve Parts.
When required
Frit – when plunger seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 ml/min of H
2
0 with purge valve opened)
Purge valve – if internally leaking
Tools required
Wrench 1/4 inch
Wrench 14 mm
Pair of tweezers or toothpick
Parts required
PFTE frit (pack of 5) 01018-22707
Purge valve G1311-60009
Valve body
Gold seal
PTFE frit
Plastic cap