66

Troubleshooting and Test Functions

Valve Fuse

Valve Fuse

Valve Fuse 0:

Channels A and B

Valve Fuse 1:

Channels C and D

The gradient valve in the quaternary pump has drawn excessive current
causing the electronic fuse to open.

Probable Causes

•

Defective gradient valve.

•

Defective connection cable (front panel to LPM board).

•

Defective LPM board.

Suggested Actions

❏ Restart the quaternary pump. If the error message appears again, exchange

the gradient valve.

❏ Exchange the connection cable.

❏ Exchange the LPM board.