Troubleshooting and Test Functions

Inlet-Valve Missing

Inlet-Valve Missing

The active-inlet valve in the quaternary pump is missing or defective.

The processor checks the presence of the active-inlet valve connector every
2 seconds. If the connector is not detected by the processor, the error
message is generated.

Probable Causes

•

Disconnected or defective cable.

•

Disconnected or defective connection cable (front panel to LPM board).

•

Defective active inlet valve.

Suggested Actions

❏ Ensure the pins of the active inlet valve connector are not damaged.

Ensure the connector is seated securely.

❏ Ensure the connection cable is seated correctly. Exchange the cable if

defective.

❏ Exchange the active inlet valve.