0 battery, Backlight, Operation – Nova-Tech GC Indicator User Manual

Page 27

Advertising
background image

P a g e

25

© Adam Equipment Company 2010

9.4

AUTO

SLEEP

FUNCTION

This function may be enabled or disabled by the user. If enabled, when the scale is
not used for some time (as pre-set by the user under this function) it automatically
switches off. To set this parameter-

• Press and hold the [Zero] key for 4 seconds, release when the display

shows “SLEEP nodE”.

• Press [U/Wt/Unit] key to scroll through the auto sleep values.

“0” Auto sleep mode disabled

“1” Auto sleep after 1 minute

“5” Auto sleep after 5 minutes

“10” Auto sleep after 10 minutes

• Press [Tare] to set the value. The scale returns to zero.

10.0 BATTERY

AND

BACKLIGHT

OPERATION

10.1 BATTERY

• The scales can be operated from the battery, if desired. The battery life

is approximately 70 hours.

• When the battery needs charging the battery symbol will appear to be

empty. The battery should be charged as possible. Once the “LO bAT”
message is shown the scale will still operate for about 10 minutes after
which it will automatically switch off to protect the battery.

• To charge the battery, simply apply power to the scale. The scale does

not need to be turned on.

• The battery should be charged for at least 12 hours for full capacity.

• Just above the “Unit Weight” display is an LED to indicate the status of

battery charging. When the scale is plugged into the mains power, the
internal battery will be charged. If the LED is green the battery is fully
charged. If it is red, the battery is nearly discharged and yellow
indicates the battery should be charged longer, preferably overnight.

• If the battery has not been used properly or it is used for a number of

years it may eventually fail to hold a full charge. If the battery life
becomes unacceptable then contact your supplier.

Advertising