12.2 USING

“2006”

TO

ENTER

THE

SERVICE

PARAMETERS

Before setting the parameters for units, decimal point, capacity and increment the
user should make sure they understand how the scale should be configured. The
scale is limited to 30,000 scale divisions. To determine the number of scale divisions
divide the capacity by the desired increment. For example a 500kg scale reading to
0.01kg would have 50,000 divisions, a value not allowed. However a 500kg scale
with interval of 0.02kg is 25,000 divisions, an acceptable number.

The order of the settings will request the default weighing unit first. This is to
assign a weighing unit to the first calibration, regardless of what he user may select
later. If the indicator has been configured for use in the USA the choice can be
grams, kilograms or pounds. For the rest of the world it is kilograms or grams only.
This setting should match the type of masses you have for calibration, imperial or
metric.

The next 2 parameters will set the decimal point position, capacity and increment. It
will be necessary for the settings to be within allowable limits for a setting to be
accepted. For example, it is only allowed that the grams can be set to 45,000 or
less. If you try to set the capacity to 50,000 grams it is not allowed, use kilograms
for the default unit and set the capacity to 50kg. When setting the interval the
options shown will only be those that allow the number of divisions to be 30,000 or
less. For example setting capacity to 5000g and the decimal point to 0.0, the
increment of 1 (=0.1g) is not allowed. Only the increment of 2, 5, 10 or 20 is
allowed.

PROCEDURE:

• Press the [Tare] key during the display counting when turned on. The

display will ask for a code number, “Pin “ on the Weight Display.

• Enter the number 2006 when “Pin “ is displayed and then press

[Tare].The displays will show the first parameters, called “F1” “CAL”.

• To select another parameter press the [U.Wt./Units] key to advance

through the parameters.

• Press the [Tare] key to enter a parameter. To exit a parameter, press

the [Zero] key.

• The Weight window will show the parameter number and the Unit

Price and Count windows will show the word describing the function.

• When a parameter is entered by pressing the [Tare] key, the displays

will guide you through the parameter selected and the options
available.