Nova-Tech GC Indicator User Manual

Page 34

Advertising
background image

P a g e

32

© Adam Equipment Company 2010

• Press the [U.Wt./Units] key to change the value. Select from 0, 0.0,

0.00, 0.000, 0.0000

• Press [Tare] to accept the displayed value.

12.2.3

F3 – CAPACITY

• To enter this parameter, press the [Tare] key when “F3” is shown. The

display will show the current capacity.

• Enter the numeric values using the keypad. The scale will check the

number of divisions n = maximum/increment is less than 30,000
divisions.

• Press [Tare] to accept the displayed value.

NOTE: The scale use load cells suitable for the capacity selected.

• The display then will go to the increment setting.

• To set the value for the increment of the scale. For example 100kg x

0.01kg the increment is 10grams. but the last digit increments by 1.

• The display will show the current increment value.

• Press the [U.Wt./Units] key to change the value. The indicator will

determine what the minimum and maximum increment is possible for
the capacity and decimal point you have set.

• Press [Tare] to accept the displayed value.

12.2.4

F4 -CALIBRATION

• To enter the calibration parameter, press the [Tare] key when “F4” is

displayed. The scale will be calibrated using 2 masses of approximately
1/3Maximum and Maximum. If the scale has been calibrated once the
values will be stored. If this is the first time the scale is calibrated the
user must enter the values for the calibration masses.

• The display will instruct you to remove any weight from the scale,

“UnloAd”. Press [Tare].

• The display will tell you to add the first weight to the scale: “LoAd 1” “

40” “kiloS”

• Add the weight shown, wait for stability then press the [Tare] key.