15

12.0 CALIBRATION

• Occasionally the scale should be verified whether it is

weighing correctly by measuring to a known mass.

• Zero the scale. Place the mass on the centre of the platform

and note the reading. Calibrate the scale, if necessary.

PROCEDURE

• While in the normal weighing mode, press and hold the

[Tare/Zero] key for 4 seconds.

• The display will show “CAL” along with the last selected unit.

The unit can be changed by using the

[Unit] key to calibrate

in kg or lb.

• Press the [Print/Hold] key. The display will show “L xx”

where xx is the Calibration weight which is user-selectable.

• Use the [Tare/Zero] key to change the flashing digit and the

[Print/Hold] key to move to the next digit.

• Press the [Unit] key to confirm the calibration weight. The

digit stops flashing.

Note: If the selected mass is less than 10% of the capacity
of the scale, an error message “CALEr” will be displayed and
the scale will return to zero. Repeat the process correctly.