Nova-Tech CPWplus User Manual

Page 24

Advertising
background image

© Adam Equipment Company 2008

22

HOLD AND PRINTING TABLE

trn 1

1

1

1

trn 2

2

2

2

trn 3

3

3

3

Hod 1

1

1

1

RS-232 is off.
Hold is off.
[Print/hold] key has
no function.

Prints continuously.
Hold is off.
[Print/hold] key has no
function.

RS-232 prints when
[Print/Hold] is pressed.
Hold function is disabled.

Hod 2

2

2

2

RS-232 is off.
Hold occurs
automatically when
the weight is stable.
Hold is released if
[Print/Hold] is
pressed or time
expires as per

Hti

setting.

Print continuously.
Hold occurs automatically
when the weight is stable.
Hold is released if
[Print/Hold] is pressed or
time expires as per

Hti

setting.

RS-232 prints and hold
occurs automatically when
the weight is stable.
[Print/Hold] key is
pressed print will occur
again.
Hold function is released
if the key is pressed again
or time expires as per

Hti

setting.

Hod 3

3

3

3

RS-232 is off
Hold occurs when the
[Print/Hold] key is
pressed.
Hold is released if
[Print/Hold] is
pressed again or time
expires as per

Hti

setting.

Print continuously.
Hold occurs when the
[Print/Hold] key is
pressed.
Hold is released if
[Print/Hold] is pressed
again or time expires as
per

Hti

setting.

RS-232 prints and the
hold occurs when
[Print/Hold] is pressed.
If

[Print/Hold] is pressed

a second time print will
occur again.
Hold is released if
[Print/Hold] is pressed
again or time expires as
per

Hti

setting.

13.2 TECHNICAL PARAMETERS





The technical parameters allow adjusting of the scale for accuracy and
speed. See section 18.0 for the complete list of parameters.

• Switch off the scale.

• Hold the [Unit] key and then press [On/Off] momentarily.

Release the

[Unit] key. The display shows the first technical

parameter to set the filter “Fi x”.

• To exit the parameter setting at any time, press the

[Print/Hold] key.

• To scroll through the technical parameters, press the [Unit]

key (which will advance to the next parameter).

• To return to normal weighing, turn the scale off and back to

on again or press the

[Print/Hold] key.