6.0  CALIBRATION 

Calibration may be required when the balance is initially installed or moved to a new 
location. Over time and use, mechanical deviations may occur so it is recommended 
that you calibrate you balance regularly for peak performance.  

Before calibrating your should make sure you have the correct weights. The weights 
should be known to an accuracy that is appropriate for the balance being calibrated, 
for  example,  OIML  Class  M1  type  or  ASTM  E617  Class  4.  If  you  do  not  have  the 
correct weights do not attempt calibration.  

Model   CQT 202 

CQT 251  

CQT 601  

CQT 1501  

CQT 2601  

CQT 2000  

CQT 5000  

Calibrat
ion 
Weight  

100g 

Or 

200g 

100g 

Or 

200g 

200g 

Or 

400g 

Or 

600g 

500g 

Or 

1000g 

Or 

1500g 

1000g 

Or 

2000g 

1000g 

Or 

2000g 

1000g 

Or 

3000g 

or 

5000g 

 

1) To enter into the calibration parameter turn the balance off, press  [Tare] and 

[Unit]  keys  at  the  same  time  and  switch  on  the  balance  after  3  seconds 
release [tare] and [unit] keys . 

2) The display will show UnLOAD remove any objects from the pan.  

3) Press the [Tare] key to enter a zero value.   

4) The  value  last  used  for  calibration  will  be  displayed  (see  above  table).  To 

select a different calibration weight, press the [Unit] key. 

5) Once  you  have  selected  the  calibration  weight  to  be  used  press  the  [Tare] 

key. 

6) LOAD will be displayed, place the calibration weight on the pan.  

7) Once the stable sign is shown press the [Tare] key.  

8) If  the  weight  is  within  5%  of  the  last  calibration  value  the  display  will  show 

PASS  and  exit  the  calibration  mode.  If  the  calibration  is  not  successful  the 
display will show FAIL and exit the calibration mode, if the calibration fails try 
it again. 