ERROR MESSAGES

During the initial power-on testing or during operation, the scale may show an error

message. The meaning of the error messages is described below.

If an error message is shown, repeat the step that caused the message. If the error

message is still shown then contact your dealer for support.

ERROR

CODE

DESCRIPTION

POSSIBLE CAUSES

Err 1

Time input Error

Invalid time entry such as “268970”

for the time format “H-m-S”.

Err 2

Date input Error

34th day of a month is an invalid

entry.

Err 4

Initial Zero is greater than

allowed (4% of maximum

capacity) when power is

turned on or when the

[Zero/Enter] key is pressed.

Weight on the platform when turning

the scale on.

Excessive weight on the platform

when zeroing the scale.

Platform is not installed.

Improper calibration of the scale.

Damaged load cell.

Damaged Electronics.

Err 6

A/D count is not correct

when turning the scale on.

Load cell is damaged.

Electronics is damaged.

Err 7

Percent input error

Percent function is entered with no

reference mass on the platform.

Err 8

High limit input error

Low limit is set first, then the high

limit is set lower than the low limit

and high limit not equal to zero.