Service parameters – Nova-Tech GFK Series User Manual

Page 53

Advertising
background image

51

|

P a g e

© Adam Equipment Company 2011

15.

SERVICE PARAMETERS

15.1.

ACCESS TO PARAMETERS

APPROVED SCALES

Access to the scale parameters and calibration is controlled in all approved scales

either by limiting access to be after the Calibration Jumper is put on the PCB,

location J1, pins 1 & 2. In this case the display will show the passcode request

screen, “ P - - - - “ . To continue enter a passcode as described below.

Or if the Calibration and Parameters have been enabled the user must enter the

correct password to have access. See Section 6.0.

Entering passcode 0000 will allow calibration as shown in 15.1, entering 1000 will

allow access to a limited set of parameters as described in section 6.2.

NON-APPROVED SCALES

Non EC Approved scales will allow entry to the parameters if the Tare key is pressed

during the power on cycle. The passwords work as above.

15.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER

“Pn”

When “Pn” is displayed.

Enter “0 0 0 0” and press [Tare]

“UnLoAd ”

Empty the platform by removing the load, if

there is any and press [Tare]

“LoAd” “6” “KiLoS”

Load the requested calibration weight and press

[Tare]

“SPAn” “PASS”

If Calibration is complete, “SPAn PASS” will be

Advertising