Apply power to the indicator. If the jumper has been used the display will ask for a

code number, “Pn “ on the Weight Display immediately. Or press the [Tare] key

during the time the calibration counters are being displayed.

Enter the number 1000 when “Pn “ is displayed and then press [Tare].

The displays will show the first parameters, called “F4” “Int”.

To select another parameter press the [Func] key to advance through the

parameters.

Press the [Tare] key to enter a parameter.

To exit a parameter, press the [Zero] key.

The display will show the parameter number and a name.

When a parameter is entered by pressing the [Tare] key, the displays will guide you

through the parameter selected and the options available.

The parameters available are:

“F4 Int”

Initial Zero Range

“F5 rEZ”

Re-Zero range

“F6 SCS”

Successive Tare Enable

“F7 Cnt”

Display the A/D counts

“F8 Zem”

Zero Mode

“F9 Lvd”

Low voltage detection

“F10 Cn”

GK-M only,

Calibration and Parameter counters or

“F10 Az”

Auto Zero Range