12

performa ce in extreme sample ranges. Use this setup function to

eter default is 1.0 to match the

n

change the cell constant if necessary. M
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).

Press

.

Press

or

to select 0.1, 1.0, or 10.0

Press

to confirm.

6.13 P4.0 rSt (Reset)

Press

.

or

Press

to select “Yes” (Reset) or “No” (Cancel).

If “Yes”, press

or

to select “Cal” (calibration reset

r “F

(complete reset to factory default settings).

only) o

Ct”

Press

to confirm.

6.14 P5.0 CLr (Clear Memory)

Press

.

Press

or

to select “Yes” (Erase memory) or “No”.

Press

to confirm.