2 single or multi-point calibration, Single or multi-point calibration – Nova-Tech CON 700 User Manual

Page 9

Advertising
background image

5

The factory default is automatic conductivity calibration. See Section
6.10

to change this setting. There is no automatic calibration value

available for conductivity range 1.

TABLE 1

Automatic Calibration Values

Normalization Temperature

Range

#

Conductivity

Range

25 ºC

20 ºC

r 1

0.00 – 20.00 μS None

None

r 2

20.1 – 200.0 μS 84

μS 76

μS

r 3

200.1 – 2000 μS 1413

μS 1278

μS

r 4

2.01 – 20.00 mS

12.88 mS

11.67 mS

r 5

20.1 – 200.0 mS

111.8 mS

102.1 mS

Range

#

TDS Range

(using 0.5 TDS factor)

Automatic Calibration

Values

r 1

0 – 10.00 ppm

none

r 2

10.1 – 100.0 ppm

none

r 3

101 – 1000 ppm

none

r 4

1.01 – 10.00 ppt

none

r 5

10.1 – 100 ppt

none

4.2 Single or Multi-Point Calibration
Use Single-Point Calibration to apply a single calibration value across
all ranges.

Use Multi-Point Calibration for individual calibration in each range. This
will restrict an individual calibration so that it is applied to one range
only. When using multi-point calibration, perform a calibration in each
range that you expect to use for best results.

Advertising