Ph mode— p1.0: viewing previous calibration data, Ph mode —p2.0: viewing electrode data – Nova-Tech pH-Dissolved Oxygen Meter User Manual

Page 17

Advertising
background image

32

33

From measurement mode:

1.

Press the MODE key to select pH
measurement mode.

2.

Press the SETUP key to enter SETUP
mode.

See figure

3.

Press the

and keys to scroll

through subgroups until you view
parameter P1.0.

4.

Press the ENTER key repeatedly to view
previous calibration data.

P1.1 = pH 1.68
P1.2 = pH 4.01
P1.3 = pH 7.00
P1.4 = pH 10.01
P1.5 = pH 12.45

See figure

5.

When you have scrolled through all
calibration data, you will automatically
return to the subgroup menu. Press the
CAL/MEAS key to return to measure-
ment mode.

NOTES:

If there is no previous calibration data at
a particular point, the primary display will
show “----”.

See figure

pH mode—P1.0: Viewing previous calibration data

Program P1.0 lets you recall previous calibration data, which helps you know when to
recalibrate your meter. This is a “view only” mode.

8.2

pH

SETUP

A

A

pH

SETUP

B

pH

SETUP

C

B

C

P2.1: Electrode offset

From measurement mode:

1.

Press the MODE key to select pH
measurement mode.

2.

Press the SETUP key to enter Set Up
mode.

3.

Press the

and keys to scroll

through subgroups until you view
parameter P2.0.

See figure

4.

Press the ENTER key to select
parameter P2.1.

5.

The display shows the electrode offset
value.
• the display shows the mV offset at

pH 7.00. If you have not calibrated at
any buffer, the primary display shows
0.00 mV.

See figure

6.

Press the ENTER key to to proceed to
step 4 of P2.2. Or, press the CAL/MEAS
key to return to measurement mode.

pH mode—P2.0: Viewing electrode data

Program P2.0 has two “view only”options that lets you check the electrode parame-
ters for diagnostic purposes.

pH

SETUP

A

A

SETUP

mV

pH

B

B

8.3