Nova-Tech pH-Dissolved Oxygen Meter User Manual

Page 19

Advertising
background image

37

36

From measurement mode:

1.

Press the MODE key to select pH
measurement mode.

2.

Press SETUP key to enter SETUP mode.

3.

Press the

▲ and ▼ keys to scroll through

subgroups until you view parameter P3.0.

See figure

4.

Press the ENTER key twice to select
parameter P3.2.

See figure

5.

Press the

▲ and ▼ keys to select 2, 3, 4,

or 5 point pH calibration.

6.

Press the ENTER key to confirm selection
and to return to the subgroup menu.
Press the CAL/MEAS key to return to
measurement mode.

pH

SETUP

SETUP

C

C

D

D

P3.2: Selecting number of pH calibration points

Program P3.2 lets you select the number of pH calibration points the meter will use
in calibration mode: 2, 3, 4, or 5. The meter will automatically exit calibration mode
after you have calibrated to your selected number of points.

P3.3 Selecting °C or °F

This meter lets you select between °C and °F units for temperature readings.

From measurement mode:

1.

Press the MODE key to select pH
measurement mode.

2.

Press SETUP key to enter SETUP mode.

3.

Press the

▲ and ▼ keys to scroll through

subgroups until you view parameter P3.0.

See figure

4.

Press the ENTER key three times to
select parameter P3.3.

See figure

5.

Press the

▲ and ▼ keys to toggle

between °C and °F.

6.

Press the ENTER key to confirm selec-
tion and to return to the subgroup menu.
Press the CAL/MEAS key to return to
measurement mode.

pH

SETUP

E

E

F

NOTES:

You can also switch between °C and °F in SETUP Program P5.2 on page 44 in
% saturation mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch
between °C and °F in pH mode, the meter will also switch in % saturation and mg/l
(ppm) modes.

C

SETUP

F

SETUP

F