Perma Pure Mini-GASS 2812T User Manual

Page 20

Advertising
background image

20

Mini-GASS User Manual

APPENDIX A

1. Specifications

MAXIMUM SAMPLE FLOW RATE:

0 TO 10 LITER/MIN

MAXIMUM INLET SAMPLE TEMPERATURE:

250

°F W/SST FILTER

230

°F W/KYNAR FILTER

MAXIMUM GAS SAMPLE WATER VAPOR CONTENT:

30%

OUTLET SAMPLE DEW POINT:

26

°F (-4°C) at 10 L/MIN

(With PD-200T-24SS dryer)

12

°F (-12°C) at 5 L/MIN

-12

°F (-25°C) at 2 L/MIN

36

°F (2°C) at 10 L/MIN

(With PD-100T-24SS dryer)

26

°F (-4°C) at 5 L/MIN

6

°F (-15°C) at 2 L/MIN

42

°F(5°C) at 10 L/MIN

(With PD-50T-24SS dryer)

34

°F(1°C) at 5 L/MIN

22

°F(-6°C) at 2 L/MIN

SOLUBLE GAS REMOVAL RATES:

NO, NO2

0% loss

SO2

0% loss

CO, CO2

0% loss

H2S, HCl

0% loss

MAXIMUM GAS SAMPLE INLET PRESSURE:

20 PSIG

MINIMUM GAS SAMPLE INLET PRESSURE:

- WITHOUT PURGE EDUCTOR OPTION:

5" H2O VACUUM

- WITH PURGE EDUCTOR OPTION:

10” H2O VACUUM

GAS SAMPLE INLET FITTINGS:

1/4" or 3/8" TUBING FITTINGS

GAS SAMPLE OUTLET FITTINGS:

1/4" or 3/8" TUBING FITTINGS

AIR REQUIREMENTS:

PURGE AIR -40

°C DEW POINT

MAXIMUM ONE (1) CFM

ELECTRICAL REQUIREMENTS:

110/220VAC, 50/60Hz, 5.0A/2.5A
(475 WATTS)

FUSE:

5 AMP BUSS TYPE AGC
OR EQUIVALENT

ENCLOSURE:

NEMA 4X, FIBERGLASS WITH
POLYCARBONATE COVER

DIMENSIONS:

- MG-1228

12”W x 28”H x 7”DEEP

- MG-2812P

28”W x 12”H x 7”DEEP

- MG-1220

12”W x 20”H x 7”DEEP

OPERATING ENVIRONMENT:

-20

°C TO 40°C AMBIENT TEMP.

0 TO 95% R.H.

Advertising