Figure 7- 1 shows the components of the L4445A microwave
switch/attenuator driver configuration. The L4445A driver is shown
connected to a single 34945EXT remote module.

L4445A Instrument Driver

Y1150A-Y1155A Distribution Board

9-pin D-SUB cable

(power to master 34945EXT)

34945EXT Extender

User-supplied switch
and cabling

Figure 7-1. L4445A Microwave Switch / Attenuator Driver Configuration.

• Each 34945EXT module can have up to four distribution boards

installed. You can have up to eight 34945EXT modules per L4445A.

• The L4445A driver interface can supply 24 V power to the first

(master) remote module only. The first remote module can also use an
external power source..

• Slave modules are connected in a daisy chain fashion using standard

ethernet RJ- 45 connectors and Cat 5 cables.

• All slave modules must obtain 24V power from an external power

supply. Each module can be powered by a separate supply.

• The Cat 5 Ethernet cable must be plugged- in to port 1 on the master

remote module. Port 1 and Port 2 are interchangeable on all slaves.

• All distribution boards on each remote module must use the same

power supply voltage.