Operating and Programming

3

L4400 User’s Guide

77

• You can specify a sweep count in conjunction with a trigger count and

a sample count. The three parameters operate independent of one
another, and the total number of readings returned will be the
product of the three parameters.

• You can store at least 500,000 readings in memory and all readings

are automatically time stamped. If memory overflows, the new
readings will overwrite the first (oldest) readings stored; the most
recent readings are always preserved.

• The

CONFigure

and

MEASure?

commands automatically set the sweep

count to 1 sweep.

• The instrument sets the sweep count to 1 after a Factory Reset

(

*RST

command). An Instrument Preset (

SYSTem:PRESet

command) or

Card Reset (

SYSTem:CPON

command) does not change the setting.

The following command sets the sweep count:

SWEep:COUNt

Channel Delay

Usage:

•

L4421A 40- Channel Armature Multiplexer

You can control the pacing of a scan sweep by inserting a delay between
the L4421A channels in the scan list (useful for high- impedance or
high- capacitance circuits). The delay occurs following relay closure and
any inherent settling time, and before the generation of the “channel
closed” signal that would externally trigger a separate DMM (see
“Scanning with External Instruments”).