• You can store at least 500,000 readings in memory and all readings

are automatically time stamped. If memory overflows, the new
readings will overwrite the first (oldest) readings stored; the most
recent readings are always preserved.

• You can specify a trigger count in conjunction with a sweep count.

The two parameters operate independent of one another, and the total
number of readings returned will be the product of the two
parameters.

• The

CONFigure

and

MEASure?

commands automatically set the scan

trigger count to 1.

• The instrument sets the scan trigger count to 1 after a Factory Reset

(

*RST

command). An Instrument Preset (

SYSTem:PRESet

command) or

Card Reset (

SYSTem:CPON

command) does not change the setting.

The following command sets the trigger count:

TRIGger:COUNt

To configure a continuous scan, send

TRIG:COUNT INFINITY

.

Sweep Count

Usage:

•

L4450A 64- Bit Digital I/O (digital input, counter channels only)

•

L4452A Multifunction Module (digital input, totalizer channels only)

The sweep count sets the number of sweeps per trigger event during a
scan (a sweep is one pass through the scan list).

Figure 3-11. Sweep Count Diagram.

• The sweep count is valid only while scanning. If no channels have

been assigned to the scan list, the specified sweep count is ignored
(no error is generated).

t

Sweep 1

Trigger

Sweep Count

(1 to 500,000 sweeps)

Sweep 2

Sweep

n