Part I: Control Unit

Part I 4-21

Teledyne Analytical Instruments

4.6.3 Precautions

The Model 2010A allows a great deal of flexibility in choosing ranges for

automatic range switching. However, there are some pitfalls that are to be
avoided.

Ranges that work well together are:

•

Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude

•

Ranges whose upper limits coincide with the lower limits of the
next higher range

•

Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.

Range schemes that are to be avoided include:

•

Ranges that overlap

•

Ranges whose limits are entirely within the span of an adjoining
range.

Figure 4-2 illustrates these schemes graphically.

Figure 4-2: Examples of Autoranging Schemes

4.7

The Analyze Function

Normally, all of the functions automatically switch back to the

Analyze

function when they have completed their assigned operations. Pressing the