Teledyne Analytical Instruments
Part II: 1-3
Thermal Conductivity Analyzer
Part II: Analysis Unit
Figure 1-2: Electrical Connector/Control Panel
•
Power In
Power input terminals for electric heater.
Requires 110 or 220 V ac, depending on
position of the Voltage Selector switch. Use
50/60 Hz.
CAUTION: Check the position of the Voltage Selector switch
BEFORE applying power to the Power Input termi-
nals.
•
Voltage Selector Power input selector switch for electric heater.
Adjusts input requirement for 110 or
220 V ac, depending on available source
voltage. Use 50/60 Hz.
•
Fuses
1.6 A, 250 V, T type, European size
5
×
20 mm fuses. Fuse 2 is on the neutral side
of the line. Fuse 1 is on the hot side of the
line.
•
Solenoid Valves Terminals that provide all electrical intercon-
nections from the Control Unit to the gas
control valves.
•
Sensor Signal
Terminals that provide connections from the
T/C sensor to the Control Unit.