Teledyne 3010MA – Paramagnetic oxygen analyzer, no 0-100% range – (superceded) User Manual
Page 37

Paramagnetic O2 Purity Analyzer
Installation
Teledyne Analytical Instruments
37
Table 3-6: Commands via RS-232 Input
Command Description
as<enter>
Immediately starts an autospan
az<enter>
Immediately starts an autozero.
co<enter>
Reports "Sensor Output" (voltage
output of the sensor itself) in VDC
For example—
Raw Sensor Output: 1.000 volts
st<enter>
Toggling input. Stops/Starts any
status message output from the RS-
232, until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-2 lists certain RS-232 values that are required by the 3010MA
implementation.
Table 3-7: Required RS-232 Options
Parameter Setting
Baud 2400
Byte 8
bits
Parity none
Stop Bits
1
Message Interval
2 seconds
3.4 Installing the Analysis Unit
Installing the Analysis unit involves:
•
Mounting the Analysis Unit
•
Making gas connections to rear panel
•
Making electrical connections to rear panel