Paramagnetic O2 Purity Analyzer

Installation

Teledyne Analytical Instruments

37

Table 3-6: Commands via RS-232 Input

Command Description

as<enter>

Immediately starts an autospan

az<enter>

Immediately starts an autozero.

co<enter>

Reports "Sensor Output" (voltage
output of the sensor itself) in VDC
For example—

Raw Sensor Output: 1.000 volts

st<enter>

Toggling input. Stops/Starts any
status message output from the RS-
232, until st<enter> is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-2 lists certain RS-232 values that are required by the 3010MA
implementation.

Table 3-7: Required RS-232 Options

Parameter Setting

Baud 2400

Byte 8

bits

Parity none

Stop Bits

1

Message Interval

2 seconds

3.4 Installing the Analysis Unit

Installing the Analysis unit involves:

Mounting the Analysis Unit

Making gas connections to rear panel

Making electrical connections to rear panel