2. To change between American and European fuses, remove the
single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.
3. Replace fuse as shown in Figu re 5-2.
4. Reassemble Housing as shown in Figure 5-1.
American Fuses
European Fuses
Figure 5-2: Installing Fuses
5.3 System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the < > arrow keys to move to More, and press
Enter.
3. Use the < > arrow keys to move to Self-Test, and press
Enter.
The following failure codes apply:
Table 5-1: Self-Test Failure Codes
Power
0 OK
1 5
V
Failure
2 15
V
Failure
3 Both
failed
Analog
0 OK
1
D/A A (0-1 V Concentration)
2
D/A B (0-1 V Range ID)
3 Both
Failed