Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual
Page 28

3 Installation
Model 3020M
3-6
Teledyne Analytical Instruments
Table 3-1: Analog Concentration Output—Example
Voltage Signal
Current Signal
% O
2
Output (V dc)
Output (mA dc)
0
0.0
4.0
1
0.1
5.6
2
0.2
7.2
3
0.3
8.8
4
0.4
10.4
5
0.5
12.0
6
0.6
13.6
7
0.7
15.2
8
0.8
16.8
9
0.9
18.4
10
1.0
20.0
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
0-25%
Non Programmable
1.00
20
NOTE:
Impendance across 4-20mA output should not exceed 1000
ohms. Note polarity of terminals when wiring to the analog
outputs. Do not short 0-1VDC output for long period of time,
or damage may occur to analyzer.
CAUTION: Voltage may be present on alarm contacts when line power
is disconnected.