Operation – Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual
Page 37

Percent Paramagnetic Oxygen Analyzer
Operation 4
4-1
Teledyne Analytical Instruments
Operation
4.1
Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
•
Establish and start an automatic calibration cycle, if desired.
(Electrically operated valves required.)
•
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
•
Calibrate the instrument.
•
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
•
Establish a security password, if desired, requiring Operator to
log in.
Before you configure your 3020M these default values are in effect:
Ranges: LO = 0-5 %, MED = 0-25 %, HI = 0-100 %.
Auto Ranging: ON
Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching.
Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 20.9 %, every 0 days at 0 hours.
Password:
T E T A I
4.2
Using the Controls
To get the proper response from these controls, turn the control toward
the desired action (ESCAPE or ENTER—DOWN or UP), and then release it.
Turn-and-release once for each action. For example, turn-and-release twice
toward UP to move the VFD screen two selections upwards on the list of
options (menu).