3 Installation
Model 3020M
3-10
Teledyne Analytical Instruments
DTR
4
Data Terminal Ready
COM
5
Common
DSR
6
Data Set Ready
RTS
7
Request to Send
CTS
8
Clear to Send
RI
9
Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•
The concentration in percent
•
The range in use (HI, MED, LO)
•
The span of the range (0-100 %, etc)
•
Which alarms—if any—are disabled (AL–x DISABLED)
•
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Four input functions using RS-232 have been implemented to date.
They are described in Table 3-4.
Table 3-4: Commands via RS-232 Input
Command
Description
as<enter>
Immediately starts an autospan.
az<enter>
Immediately starts an autozero.
st<enter>
Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-5 lists certain RS-232 values that are required by the 3020M imple-
mentation.
Table 3-5: Required RS-232 Options
Parameter
Setting
Baud
2400
Byte
8 bits
Parity
none
Stop Bits
1
Message Interval
2 seconds