Teledyne 3110 – Portable Trace/Percent Oxygen Analyzer, FM-approved configuration User Manual
Page 32

Operation Model
3110
Teledyne
Analytical
Instruments
24
1. Press ENTER to stop the data logger and return to the LOG
RESET & START screen.
3.2.11 LOG TRANSMIT Screen
Data can be downloaded to a computer using the 9-pin D-
subminiature port and cable attached to a PC. The D-sub port accepts a
standard data cable with a DB-9 connector. The computer must be able
to accept data from a source with the following characteristics:
9600 baud
8 bit data
no parity
1 stop bit
no flow control
Refer to your computer manual for details on how to setup the RS-
232 communications port on your computer.
To download a data log:
1. Use the UP/DOWN keys to navigate to the LOG TRANSMIT
screen. Then press ENTER to enter the setup screen.
2. Press ENTER again to transmit the current data log.
Note: The data set will be transmitted each time you press the
ENTER key. Use the ESC key to exit the LOG TRANSMIT