Appendix – Teledyne 3110 – Portable Trace/Percent Oxygen Analyzer, FM-approved configuration User Manual
Page 43

Portable Trace Oxygen Analyzer
Appendix
Teledyne
Analytical
Instruments
35
Appendix
A.1 Specifications
Certification: FM-approved, Intrinsically Safe,
Class I Div 1 Group A-D T6
Micro-fuel Cell Class: B-2C, B-1, A-2C, A-5, B-2CXL, A-
2CXL, B-2C Insta-trace, A-2C Insta-trace,
B-2CHS, A-2CNG
Ranges of Analysis: 1 user defined range
0-10 ppm to 0-25% (std 3110)
0-2 ppm to 0-25% (3110XL)
0-1% to 0-25% (3110P)
Accuracy: ±2% of scale or ±1 PPM (whichever is
greater) at constant temperature;
±5% of reading or ±1 PPM (whichever is
greater) over the operating temperature
range.
Operating Temp. Range: 0°C to 40°C (32°F to 104°F)
Sample Temp. Range: 0-40°C
Sample Flow Rate: 0.5 – 2.5 SCFH
Recommended Span Gas: Atmospheric air for percent analysis or
80-90% of range most likely to be used
Signal Output: 0-1
VDC
Electrical safety parameters:
Uo = 4.65 V, Io = 193.74 mA,
Po = 225.2 mW, Co = 90 uF,
Lo = 0.7 mH, Lo/Ro = 157.86 uH/Ohm
Dimensions: 10.9” x 6.2” x 5.2” deep
Weight: 4.4 lbs. (2.0 kg.)
Power Requirements: 100-240 VAC 47-63 Hz Charging Adapter