Maintenance

Teledyne

Analytical

Instruments

31

WARNING:

EVIDENCE OF DAMAGE DUE TO TAMPERING OR
MISHANDLING WILL RENDER THE CELL
WARRANTY NULL AND VOID.

4.7 Temperature Compensation

The micro-fuel cell has an inherent positive temperature coefficient.

Compensation is performed internally by the microprocessor and needs no
further adjustment.

Contact the factory if you suspect temperature related inaccuracies

during monitoring and are operating within the instrument’s temperature
operation range.

4.8 Leak Testing

WARNING:

IF A LEAK IS SUSPECTED IN THE UNIT, DO NOT
ATTEMPT TO TIGHTEN THE DISCONNECT
FITTINGS. THE FITTINGS ARE POTTED IN EPOXY
AND TIGHTENING THEM WILL BREAK THE SEAL.

To check for leaks, TAI recommends one of the following

procedures:

Procedure I:

1. Purge the instrument to as low a value as possible. Use a

sensitive range for analysis for instance, 0-1% range Take
note of the oxygen concentration.

2. Place the vent line in water and disconnect the sample.

3. Next, disconnect the vent line and use the same analysis

range. The reading should not increase above the level to
which you purged to.

Procedure II:

1. Purge the instrument with nitrogen at the sample port.

2. Note the reading once is has stabilized (at least 24 hours on

the 0-10 PPM range) and at a flowrate of approximately 1
SCFH.

3. Increase the flowrate from 1 SCFH to 2 SCFH.