Appendix – Teledyne 6700-C Total Organic Carbon Analyzer (Hi-Temp Combustion version) User Manual
Page 57

Percent Oxygen Analyzer
Appendix
Teledyne
Analytical
Instruments
57
Appendix
A.1 Specifications
(The following specifications are nominal and depend on the actual
application).
Measuring
Method:
TC
(Total Carbon) – High Temperature
Combustion.
TOC (Total Organic Carbon) acidification
and sparging to eliminate inorganic carbon
interference.
Range: 0-100 through 0-10,000 ppm, full scale.
(Range changes often require carrier gas
adjustments. This analyzer uses the
precision of computer controlled mass
flow controllers to eliminate operator error
and the use of inaccurate mechanical flow
meters.)
Display: Windows with Paperless Chart Recorder
Data Handling: RS-232C, RS-485
Output:
4-20
mA
Alarms: Two alarm levels
One master fault alarm
Response
Time:
Application
dependent
Repeatability: ±3% FS (std. dev. 1 sigma)
Zero/Span Stability: ±3% FS
Linearity:
±3%
FS
Inlet Pressure: Atmospheric to 3 psig ±0.5 psi
Carrier Gas Flow Rate: 300 ccm (max.)