Teledyne BDS-3000 - Oxygen analyzer User Manual
Page 20
Introduction
BDS 3000
Teledyne Analytical Instruments
6
•
Alarm Connections
2 concentration alarms and 1
system alarm.
•
RS-232 Port
Serial digital concentration signal
output and control input.
•
Remote Probe
Used in the BDS 3000 for
controlling external solenoid
valves only.
•
Remote Span/Zero
Digital inputs allow external
control of analyzer calibration.
•
Calibration Contact
To notify external equipment that
instrument is being calibrated and
readings are not monitoring
sample.
•
Range ID Contacts
Four separate, dedicated, range
relay contacts. Low, Medium,
High, Cal.
•
Network I/O
Serial digital communications for
local network access. For future
expansion. Not implemented at this
printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model BDS 3000 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.