Installation

BDS 3000

Teledyne Analytical Instruments

24

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 V dc

23

– Range ID, 0-1 V dc, negative ground

24

+ % Range, 0-1 V dc

7

– % Range, 0-1 V dc, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:

Threshold Alarm 1:

•

Can be configured as high (actuates when concentration

is above threshold), or low (actuates when concentration
is below threshold).

•

Can be configured as failsafe or nonfailsafe.

•

Can be configured as latching or nonlatching.

•

Can be configured out (defeated).

Threshold Alarm 2:

•

Can be configured as high (actuates when concentration

is above threshold), or low (actuates when concentration
is below threshold).

•

Can be configured as failsafe or nonfailsafe.

•

Can be configured as latching or nonlatching.

•

Can be configured out (defeated).