Relay pca status leds – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 271

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

271

12.3.2.2. Relay PCA Status LEDs

D7 (Green) – Optional Zero/Span Valve

D11 (Green) – Optional Dual Span Select Valve

D12 (Green) – Optional Pressurized Span Shutoff Valve

D3 (Yellow) NO

2

NO Converter Heater

D5 (Yellow) – Optional Internal Span Gas Gen Heater

D8 (Green) – Optional Sample/Cal Valve

D1 (RED)

Watchdog Indicator

D2 (Yellow) Reaction Cell Heater

D9 (Green) – AutoZero Valve

D10 (Green) – NO/NO

x

Valve

D13 (Green) – Optional Pressurized Zero Shutoff Valve

Figure 12-3:

Relay PCA Status LEDS Used for Troubleshooting

Table 12-5: Relay PCA Status LED Failure Indications

LED COLOR

FUNCTION

FAULT

STATUS

INDICATED FAILURE(S)

LED ROW 1

D2

Yellow

Reaction Cell heater

Continuously

ON or OFF

Heater broken, thermistor broken

D3 Yellow NO

2

converter heater

Continuously

ON or OFF

Heater broken, thermocouple broken

D7

Green

Zero/Span valve status

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D8

Green

Sample/Cal valve status

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D9

Green

Auto-zero valve status

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D10 Green

NO/NO

x

valve status

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

LED ROW 2

D5 Yellow

Internal span gas generator

perm tube heater

Continuously

ON or OFF

Heater broken, thermistor broken

D11

Green

Dual span select valve

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D12

Green

Pressurized Span shutoff valve

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D13

Green

Pressurized Zero shutoff valve

Continuously

ON or OFF

Valve broken or stuck, valve driver chip broken

D14- 16

Green

Spare

N/A N/A