Gas conditioner options, The 9110t includes a nafion, Permeation gas exchange tube to remove h – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 72

Model 9110T NOx Analyzer
Getting Started
Teledyne Analytical Instruments
72
Zero/Span
Open to ZERO AIR inlet
3 2
ZERO CAL
Sample/Cal
Open to ZERO/SPAN Valve
1 2
Zero/Span
Open to ZERO AIR inlet
3 2
SPAN CAL
Sample/Cal
Open to ZERO/SPAN Valve
1 2
Zero/Span
Open to SPAN inlet
1 2
3.3.2.6. Gas Conditioner Options
AMMONIA REMOVAL SAMPLE CONDITIONER (OPT 86A)
The 9110T includes a Nafion
®
permeation gas exchange tube to remove H
2
O a from the ozone generator
supply gas stream to a dew point of about -20° C (~600 ppm H
2
O) and effectively remove concentrations
of ammonia (NH
3
) up to about 1 ppm.
An additional Sample Conditioner can be added to the 9110T’s sample gas stream.
SAMPLE
PRESSURE
SENSOR
VACUUM
PRESSURE
SENSOR
O
3
FLOW
SENSOR
PMT
PERMAPU
RE
DRYE
R
EXH
AUST MANIFO
LD
NO/NO
X
VALVE
AUTOZERO
VALVE
Sample Gas
NH
3
Dryer/Scrubber
Figure 3-28:
Pneumatics for Sample Conditioner OPT 86A
ZERO AIR SCRUBBER (OPT 86C), FOR Z/S VALVES